A LC-MS/MS method is a reference method of determining a contaminant
in food. An
ELISAmethod
was developed for screening the contaminant. The method validation of
ELISA was composed of two parts, the first part was to verify whether
the ELISA method is fit for the screening purpose, the second part was
to verify whether the ELISA method is comparable with the LC-MS/MS
method.
The analysis of validation data revealed that the ELISA method is accurate and precious and it could be used for screening purpose.
This work demonstrates the author’s skills in statistical analysis in R and reporting with R Markdown. I only provided some of method validation results in this work.
Part I: Method validation of ELISA
The precision and accuracy of the ELISA were evaluated using replicate
fortified samples (n = 5 replicates per day over three days).
1) A blank sample was grounded followed by being fortified at 1, 5, and 10 ppm. Five replicate test portions at each fortified level were tested every day for three days.
2) A blank sample, control standard at 0.75ppb, and QC sample were analyzed in every run. All extract from each test portion, standard, and QC sample were analyzed in duplicate.
3) Duplicates within 20% were considered valid and were calculated to obtain results; otherwise, extraction and preparation were repeated.
4) When fortified samples initially produced results outside of the range of the calibration curve, additional dilutions were performed on the extract to allow for analysis on the curve. Final dilutions of 1:200, 1:400, and 1:600 (v/v) were used when appropriate.
Part II: Comparison of ELISA with LC-MS/MS
30 samples were tested on ELISA plate as well as with LC-MS/MS.
Precision is defined as the closeness of agreement between independent test results obtained under stipulated conditions. [1]
Precision was evaluated by assessing the mean and the associated
standard deviation at different fortification levels, as well as the
within-day
%CV
[2] and between-day %CVs. In this method
validation study, I followed these criteria: The within-day %CVs should
be less than 10. The between-day %CVs of less than 15 are generally
acceptable [3].
Accuracy was evaluated by assessing the percent recovery of fortified samples at different fortification levels.
| Stats | Day 1 | Day 2 | Day 3 |
|---|---|---|---|
| Sample fortified at 1ppm | |||
| mean ± SD | 0.98 ± 0.15 | 0.94 ± 0.10 | 1.04 ± 0.16 |
| mean % recovery | 98 | 94 | 104 |
| %CV | 14.79 | 10.38 | 15.07 |
| Sample fortified at 5ppm | |||
| mean ± SD | 4.44 ± 0.41 | 6.00 ± 0.41 | 5.24 ± 0.28 |
| mean % recovery | 89 | 120 | 105 |
| %CV | 9.21 | 6.86 | 5.4 |
| Sample fortified at 10ppm | |||
| mean ± SD | 10.68 ± 0.84 | 10.02 ± 1.00 | 10.51 ± 0.65 |
| mean % recovery | 107 | 100 | 105 |
| %CV | 7.88 | 9.95 | 6.18 |
Table II summarized the within-day precision and accuracy. Figure II and III visualized the precision and accuracy.
The ELISA method had good accuracy, with the
mean % recovery ranging from 89 to 120%, at each
fortification level (1, 5, 10 ppm) in each day (table II).
With the %CVs ranging from 5.4 to 9.95%, the ELISA method had good precision at 5 and 10ppm fortification levels in each day, but the %CVs were between 10.38 and 15.07% at 1 ppm fortification levels.
| Stat | 1 ppm | 5 ppm | 10 ppm |
|---|---|---|---|
| %CV | 4.65 | 8.46 | 3.33 |
The overall between-day precision is good with %CV ranging from 3.33 to 8.46%.
The performance of the ELISA was further evaluated by comparison of
ELISA-generated data to those from the LC-MS/MS method. The raw data
less than the LOQs were altered by removing the < symbol
for plotting purpose.
| ELISA | LCMSMS |
|---|---|
| 0.32 | <0.05 |
| 0.54 | 0.302 |
| 4.41 | 3.64 |
| 1.81 | 1.678 |
| 2.95 | 2.684 |
| <0.30 | 0.114 |
| 0.74 | 0.617 |
| 0.93 | 0.649 |
| 0.39 | 0.282 |
| <0.30 | <0.05 |
| 1.75 | 1.24 |
| <0.30 | 0.057 |
| 0.67 | 0.352 |
| 0.74 | 0.45 |
| <0.30 | 0.061 |
| 7.12 | 6.702 |
| 7.26 | 7.497 |
| <0.30 | <0.05 |
| 0.62 | 0.389 |
| 0.3 | <0.05 |
| 0.54 | 0.502 |
| 1.08 | 0.865 |
| <0.30 | <0.05 |
| 1.44 | 1.517 |
| <0.30 | <0.05 |
| 8.67 | 8.463 |
| 1.11 | 0.889 |
| <0.30 | <0.05 |
| 6.13 | 5.605 |
| <0.30 | 0.169 |
There was good correlation between the ELISA and LC-MS/MS results. Using the equation that describes the relationship between the ELISA and LC-MS/MS results, the ELISA results within the range from LOQ to 9 ppm were approximately 0.99 times lower than the LC-MS/MS results.
The ELISA test kit was validated. The accuracy in the concentration
range of 1 - 10 ppm was good. The within-day precision at 5 and 10 ppm
were good; the within-day precision at 1 ppm was poor with %CVs over
10%. The between-day precision in the range of 1 - 10 ppm was
good.
The ELISA produced identical results comparing with the LC-MS/MS
method. Overall, the ELISA is a useful tool that is complemented by the
comprehensive LC-MS/MS method.
The mean in table II is calculated with the following
equation (1). The mean of each replicate was aggregated by
fortification level and day for calculation.
\[\begin{equation}
\tag{1}
mean_{conc.,day} = \frac{\sum\limits_{i=1}^{n}[mean\: of\:
duplicates]_i}{n}
\end{equation}\]
The
sd
in table II is calculated with the following equation (2). The
sd of each replicate was aggregated by fortification level
and day for calculation.
\[\begin{equation} \tag{2} sd_{conc.,day} = \sqrt{\frac{\sum\limits_{i=1}^{n} ([mean\: of\: duplicates]_i - \overline{[mean\: of\: duplicates]})^{2}} {n-1}} \end{equation}\]
The mean % recovery in table II is calculated with the
following equation (3). The mean % recovery of each
replicate was aggregated by fortification level and day for
calculation.
\[\begin{equation} \tag{3} mean\: \%\: recovery_{conc., day} = \frac{\sum\limits_{i=1}^{n}[mean\: \%\: recovery]_i}{n} \end{equation}\]
The %CV in table II is calculated with the following
equation (4). The %CV of each replicate was aggregated by
fortification level and day for calculation.
\[\begin{equation}
\tag{4}
\%CV = \frac{sd_{conc., day}}{mean_{conc.,day}} \times 100
\end{equation}\]
The %CV in table III is calculated with the following
equation (5). The %CV of each replicate was aggregated by
fortification level for calculation.
\[\begin{equation} \tag{5} \%CV_{conc.} = \frac{\sum\limits_{i=1}^{m}[\%CV]_i}{m} \end{equation}\]
[1] ISO 5725-1. Accuracy (Trueness and Precision) of Measurement Methods and Results – Part1: General Principles and Definitions (1994).
[2] %CV in ELISA: How to Reduce Them and Why They’re Important
[3] Calculating Inter- and Intra-Assay Coefficients of Variability
[4] Andreasson, U., Perret-Liaudet, A., van Waalwijk van Doorn, L. J. C., Blennow, K., Chiasserini, D., Engelborghs, S., … Teunissen, C. E. (2015). A Practical Guide to Immunoassay Method Validation. Frontiers in Neurology, 6. doi:10.3389/fneur.2015.00179
Created Date: 2022-08-30
Last Modified Date: 2022-10-03