Sonde at IBWC gage

Serial number 12215259

1. Table of QAQC notes.

Table created using field notes and visual interpretation of time series. Notes file is Visual examination performed on wqdatalive and using plot_ly in 2_interactive_qaqc_each_parm.R.

X = parameter invalid (sensor dry, sedimentation, malfunction) CAL = calibration problem; fixable by back-calibrating to RV values VAR = Values erratic or variable; potential issue with quality RNG = values out of range of the sensor

Start End Parameter Screen.label Comment.short Comment.long
10/14/2022 19:30 11/8/2022 12:30 ALL X Sonde dry Sonde.dry
11/9/2022 14:20 11/15/2022 10:00 ALL X Sonde buried Sonde buried; cleaned by SG on 2022.11.15
10/15/2022 12:00 11/14/2022 18:10 TRYP CAL TRYP cal off TRYP calibration off by 10x; use RV and standard calibration
10/15/2022 12:00 11/14/2022 18:10 Turb.Manta CAL TURB.Manta cal off TURB Manta calibration error; cal log shows negative values for 100 NTU. Use RV and standard calibration
11/8/2022 19:10 11/9/2022 0:30 Turb.SS X TURB.SS.TU error TURB OBS SS low during peak of storm. Possible occlusion of the sensor
11/8/2022 19:10 11/9/2022 0:30 Turb.BS X TURB.BS.TU error TURB OBS BS low during peak of storm. Possible occlusion of the sensor
11/9/2022 0:40 11/9/2022 5:30 Turb.SS VAR TURB.SS.TU.variable TURB OBS SS high and variable. May need additional smooothing
11/9/2022 0:40 11/9/2022 5:30 Turb.BS VAR TURB.BS.TU.variable TURB OBS BS high and variable. May need additional smooothing
11/8/2022 17:10 11/9/2022 3:00 Turb.Manta RNG TURB.Manta out of range TURB.Manta out of range
11/9/2022 11:19 11/9/2022 11:21 Turb.Manta RNG TURB.Manta out of range TURB.Manta out of range

Example of interactive plot used to determine QAQC break dates and times, here for Turbidity OBS side-scatter.

2. Time series plot, entire period

Plot of all raw data (black line), and final raw data (blue). Gray boxes indicate periods where all parameters were not valid. This can be caused by the sonde being dry, by significant biofouling, or by sedimentation on the sonde.

For November 14, 2022 download, the sonde was dry until Nov 8, and had significant sedimentation on Nov 9:

3. Time series plot, period with valid data

2022-11-14 notes: Most parameters had valid values over the period. TRYP and CDOM were low due to miscalibration; the raw “return values” were therefore back-calculated using the calibration logs, and a second value of TRYP and CDOM calculated using the long-term stable calibration values for turning RV into ppb. Plots below are the raw values prior to recalibration.

Next: recalculate TRYP and CDOM using default calibrations. Calculate % ww using the TRYP-WW regression from the WW lab addition experiments.

4. Time series of turbidity, all turbidimeters

Turbidity values followed the same pattern for all 3 sensors but had some unlikely value drops at the storm peak. The turbidity sensors then read reasonable values on the falling limb, until burial.

5. Time series plot, including quality flags

VAR = variable, possible quality problems RNG = variable out of range CAL = requires recalibration X = variable invalid

5.1. TRYP and CDOM with flags

TRYP requires recalculation with different calibration parameters–should be higher than that for the raw sewage present at the beginning of the storm.

5.2 Turbidity plots with flags