Multicenter Test-Retest and Acquisition Protocol Reproducibility of Liver Iron Quantification - Correlation with Ferriscan-LIC

Author

Lu Mao

LIC vs R2*

The R-sqr of linear fit between LIC and R2* are tabulated in Table 1 below by field strength and protocol (left/right lobe and overall). P01 & P02 generally have better fit than the rest.

Table 1: R-squared of linear fit between LIC and R2* by field strength and protocol.
Field Protocol Left Right All
1.5T P01 0.892 0.899 0.899
P02 0.898 0.902 0.905
P03 0.731 0.864 0.807
P04 0.801 0.877 0.848
P05 0.736 0.812 0.784
P06 0.853 0.898 0.885
P07 0.720 0.777 0.750
P08 0.758 0.815 0.795
P09 0.623 0.723 0.682
P10 0.610 0.649 0.632
3.0T P01 0.876 0.903 0.894
P02 0.859 0.888 0.877
P03 0.476 0.601 0.554
P04 0.480 0.672 0.595
P05 0.279 0.420 0.360
P06 0.246 0.390 0.328
P07 0.392 0.604 0.524
P08 0.356 0.533 0.460
P09 0.607 0.785 0.708
P10 0.303 0.423 0.380
Figure 1: Regression lines of Ferriscan LIC vs R2* (overall).