Phantoms

The relationships between Z-number and artifact length, width, intensity, and needle width are plotted below. All except artifact length are significantly associated with Z-number. In particular, the relationship between needle width and Z-number appears to be non-monotone and is fitted by a quadratic curve (a linear fit yields low \(R^2\) and non-significant \(p\)-values).

Swine model

In the swine model, artifact length, width, intensity, and needle width are compared between aluminum and stainless by kidney, liver, and lung below. The variables are summarized by median (IQR), with \(p\)-values computed by the Wilcoxon rank sum test.

Table 1. Comparison between aluminum vs stainless.
Aluminum Stainless p-value
Artifact Length (mm) - Kidney 2.8 (2.8, 2.8) 81.6 (52, 96.3) 0.004
Artifact Length (mm) - Liver 2.8 (2.8, 15.5) 92.8 (89.5, 97.7) 0.007
Artifact Length (mm) - Lung 28.8 (26.2, 47.8) 116.4 (110.9, 131.3) 0.065
Artifact Width (mm) - Kidney 0.7 (0, 2.3) 9.8 (7, 10.8) 0.008
Artifact Width (mm) - Liver 1.5 (0.4, 2.3) 9.8 (8.2, 11.2) 0.005
Artifact Width (mm) - Lung 3.4 (3.2, 4.3) 8.9 (7.8, 10.6) 0.065
Artifact Intensity (HU) - Kidney -126.2 (-215.9, -80.9) -464.4 (-666.6, -405.6) 0.002
Artifact Intensity (HU) - Liver -35.1 (-40.7, -19.6) -698.7 (-806.3, -538) 0.002
Artifact Intensity (HU) - Lung -88.3 (-110.8, -73.4) -562 (-728.2, -496.1) 0.002
Needle Width (mm) - Kidney 4.1 (2.4, 19.9) 25.7 (23.3, 27.6) 0.180
Needle Width (mm) - Liver 2.3 (1.8, 2.7) 26.1 (24.8, 27.2) 0.002
Needle Width (mm) - Lung 3.1 (3, 3.8) 25.5 (25.2, 25.7) 0.002

The following shows the bar charts comparing the means of aluminum and stainless by kidney, liver, and lung (with error bars indicating the standard error of the mean; * \(p<0.05\), ** \(p<0.01\)).